Tecan Freedom EVO with Touch Tools Suite Version 3.0 The intended use of the Touch Tools Suite is to run previously defi
FDA device recall Z-1560-2013 · posted 2013-06-19 · Terminated
Recall details
- Recalling firm
- Tecan US, Inc.
- Reason for recall
- WIth TouchTools 3.0 after the execution of a Rich User Prompt in Display and Wait mode, door lock errors are ignored. If an operator presses "Continue" following a Rich User Prompt and the door has been left open, the script will progress (continue without warning). This could lead to the instrument moving unexpectedly and potentially cause an injury to the operator if they are physically inside t
- Action taken
- The firm, Tecan, sent an "Urgent Field Corrective Action" letter dated February 15, 2013 to its consignees/customers and also contact them by phone February 19, 2013. The letter describes the product, problem and actions to be taken. The customers were instructed to not use Rich User Prompts in Display and Wait mode until TouchTools 3.0 Patch 1 or a higher version has been installed. Tecan is currently testing and releasing a patch that will correct this issue. The patch will be available by the end of February. Tecan will contact you upon its release as installation of the patch is required. If you have any further questions, please contact your local Tecan Helpdesk for assistance or call 919-361-5200 ext 19519.
- Root cause
- Software change control
- Status
- Terminated
- Product code
- JQW
- Quantity affected
- 12 units
- Distribution
- Worldwide distribution: US (nationwide) including states of: CA, KY, NC, OH, TX, and WI; and countries of: France, Germany and Italy.
- Date initiated
- 2013-02-19
- Date posted
- 2013-06-19
- Date terminated
- 2013-06-20
- Location
- Morrisville, NC
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.