TDWorkstation version 11.01.A or higher. TDWorkstation Evolution. TDWorkstation receives patient results from a connecte
FDA device recall Z-1310-2014 · posted 2014-03-31 · Terminated
Recall details
- Recalling firm
- Technidata S.A.
- Reason for recall
- If a conflict occurs between a Demography Query and a Result Transmission by the instrument, the TDWorkstation could send an incorrect patient demography to an instrument.
- Action taken
- An email dated 2/17/14 was sent to their customers to inform them about concerns only if their site uses TDWorkstation and TD-Workstation Evolution products that meet all the listed criterias. The criterias are listed in the customer notification letter and it informs the customers of the problems identified and the actions to be taken. Customers are instructed to complete the reply form and return it by fax (514) 270-2770 with in 15 days from receiving the customer notification form. Customers that do not respond the reply form, Technidata will consider those customers do not wish to install the correction.
- Root cause
- Software design
- Status
- Terminated
- Product code
- JQP
- Quantity affected
- 85 units total (1 unit in US)
- Distribution
- Distributed in the state of PA.
- Date initiated
- 2014-02-17
- Date posted
- 2014-03-31
- Date terminated
- 2014-03-31
- Location
- Montbonnot St Martin, France