TDHistology/Cytology Classification name: Calculator/data processing module for clinical use (862.2100). Intended use: L

FDA device recall Z-1743-2013 · posted 2013-07-17 · Terminated

Recall details

Recalling firm
Technidata S.A.
Reason for recall
A recall was initiated by Technidata because use of symbols (less than or equal to or greater than or equal to or any other symbol) in Microsoft Word reports when using electronic transmissions of results by of HPRIM (including HPRIM Medecin via Kermit) or HL7 must be absolutely prohibited. Such symbols are not supported by the Microsoft Word interface used by the Histology/Cytology software and
Action taken
TD Technidata Medical Software sent a Medical Device Recall letter dated June 20, 2013, to one affected customer who received the TD-Synergy Histology/Cytology module. The customer was informed of the problems identified and the actions to be taken. Customers were instructed to complete the attached Reply Form and return it to the firm by fax to 514-270.2770. The customer were also instructed to contact Technidata Canada at (514) 270-7777 or at support-canada@technidata-web.com.
Root cause
Software design
Status
Terminated
Product code
JQP
Quantity affected
1 site
Distribution
US Distribution CA only.
Date initiated
2013-06-20
Date posted
2013-07-17
Date terminated
2013-07-29
Location
Montbonnot St Martin, France