Tarsys 1st Generation Seating System - an optional accessory for the Powered Wheelchairs.

FDA device recall Z-0682-03 · posted 2003-03-26 · Terminated

Recall details

Recalling firm
Invacare Corp
Reason for recall
A short circuit within the charger harness may cause heat damage to the units with potential for fire.
Action taken
On 3/10/2003, the firm mailed recall letters to all affected dealer accounts. The dealers were asked to contact all customers who have purchased powered wheelchairs with the optional Tarsys Seating System.
Root cause
Other
Status
Terminated
Product code
ITI
Quantity affected
20,956 units
Distribution
The scooters were shipped to dealerships located nationwide, and to foreign distributors located in Thailand, So. Africa, Hong Kong, New Zealand, Japan, Maylasia, Korea, Isreal, Australia. 92% of affe
Date initiated
2003-03-10
Date posted
2003-03-26
Date terminated
2012-04-13
Location
Elyria, OH