Target Detachable Coils: TARGET HELICAL NANO 2MM X 6CM MODEL Number:M0035452060 Neurology: Target Detachable Coils are i

FDA device recall Z-2181-2016 · posted 2016-07-11 · Terminated

Recall details

Recalling firm
Stryker Neurovascular
Reason for recall
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.
Action taken
Stryker Neurovascular sent an Urgent Medical Device Voluntary Recall letter dated June 3, 2016, to all affected customers via FED Ex. Letter includes recall letter, list of impacted lot numbers, customer acknowledgement form and frequently asked questions. Customers were instructed to immediately check their inventory. segregate the effected inventory in a secure location for return to Stryker, circulate the Field Safety Notice internally to all interested/affected parties. Inform Stryker if any of the subject devices have been distributed to other organizations. Customers were also instructed to complete the attached customer response form and return the completed form to their nominated Stryker Representative.
Root cause
Process change control
Status
Terminated
Product code
HCG
Quantity affected
45,302 (8,188 in the U.S.) - total, all model numbers
Distribution
Worldwide Distribution - US (nationwide) and Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Ge
Date initiated
2016-06-06
Date posted
2016-07-11
Date terminated
2017-03-03
Location
Fremont, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
Target Detachable Coils (K153658)Stryker Neurovascular2016-05-19