Atlas FDA

Tapered Screw-Vent Implant, MTX, 4.7mm x 11.5mm, 4.5mm Platform. Catalog Number TSVWB11, Lot 62470008. Class II, 510(k)

FDA device recall Z-0847-2014 · posted 2014-01-28 · Terminated

Recall details

Recalling firm
Zimmer Dental Inc
Reason for recall
Zimmer Dental is conducting a voluntary recall of a single lot of the Tapered Screw-Vent Implant, Catalog TSVWB11, Lot 62470008, because some of the packages of this lot may have the Cap Label state 3.7 x 10mm instead of 4.7 x 11.5mm.
Action taken
Zimmer Dental initiated the recall on 01/06/2013 by sending out recall notifications via FedEx to domestic customers and via electric mail to international customers. Customers are sent a letter identifying the issue, product description with codes, risk, instructions, and contact information. The letter's instructions include locating the affected product and return unused product to Zimmer Dental and notifying the amount of product that has been consumed.
Root cause
Labeling Change Control
Status
Terminated
Product code
DZE
Quantity affected
307
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Thailand, Japan, Hong Kong, Morocco, and UAE.
Date initiated
2014-01-06
Date posted
2014-01-28
Date terminated
2014-07-07
Location
Carlsbad, CA