System 1 base 220/240V The Terumo Advanced Perfusion System 1 is indicated for use in extracorporeal circulation of bloo
FDA device recall Z-2178-2012 · posted 2012-08-09 · Terminated
Recall details
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Reason for recall
- Terumo Cardiovascular Systems sent a Safety Advisory (AA-2010-007-C) on December 2, 2010 (Recall# Z-0882/0883-2011 RES 57416) regarding malfunctions of the Occluder and Air Bubble Detection (ABD) Modules to a confined number of Terumo Advanced Perfusion System 1 users. The investigation of the malfunction found a faulty component on the modules' application board caused the failures. Terumo CVS re
- Action taken
- The firm, Terumo CVS, sent an "URGENT MEDICAL DEVICE RECALL" letter dated July 26, 2012 to its customers. The letter described the product, problem and actions to be taken. To customers were instructed to review this Medical Device Recall notice and addendum; assure that all users are aware of this notice and addendum; place the addendum in the front of Section 7: Modules (p 7.1) and confirm receipt of this communication by faxing the attached Customer Response Form to the fax number/email address indicated on form. Terumo CVS is issuing a Safety Advisory to alert all Terumo System 1 users of the potential for failure and providing an addendum for the Operator Manual with instruction on how to respond to the failure. If you have any question or concerns contact Terumo CVS Customer Service at 1-800-521-2818. Customer Service hours are Monday-Friday, 8AM - 6PM ET.
- Root cause
- Component design/selection
- Status
- Terminated
- Product code
- DTQ
- Quantity affected
- 649 units
- Distribution
- Worldwide distribution: USA (nationwide) and countries of: ARGENTINA, AUSTRALIA, BELGIUM, CANADA, CHILE, COLOMBIA, Costa Rica, Dominican Republic, Egypt, Germany, Guatemala, Honduras, Hong Kong, India
- Date initiated
- 2012-07-26
- Date posted
- 2012-08-09
- Date terminated
- 2012-11-16
- Location
- Ann Arbor, MI