System 1 Base 100/120V The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorpore
FDA device recall Z-1970-2012 · posted 2012-07-09 · Terminated
Recall details
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Reason for recall
- Terumo Cardiovascular Systems (Terumo CVS) has received reports of various roller pump malfunctions of the Terumo Advanced Perfusion System 1 that could result in the inability to use the pump and would then require replacement with back-up equipment. Description of malfunctions: " Unplanned pump stop or pause " Changes in pump speed If it is necessary to replace a roller pump during cardiopu
- Action taken
- Terumo CVS sent an "URGENT MEDICAL DEVICE RECALL" letter dated May 4, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. Customers were asked to complete and return a Customer Response Form to the firm via fax at 1-800-292-6551. Contact the firm at 1-800-521-2818 for questions regarding this recall.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- DTQ
- Quantity affected
- 992 units
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of ARGENTINA, AUSTRALIA, BELGIUM, CANADA, CHILE, COLOMBIA, COLOMBIA, Costa Rica, Dominican Republic, Egypt, Germany, Guatamala, Honduras, Hong
- Date initiated
- 2012-06-25
- Date posted
- 2012-07-09
- Date terminated
- 2013-03-01
- Location
- Ann Arbor, MI