Atlas FDA

Sysmex, Automated Hematology Analyzer XN Series, Sysmex Corporation, 1-5-1 Wakinohama-Kaigandori, Chuo-ku, Kobe 651-0073

FDA device recall Z-2034-2017 · posted 2017-04-14 · Terminated

Recall details

Recalling firm
Sysmex America Inc
Reason for recall
Reported customer complaints of automated hematology analyzers catching fire due to misalignment of an internal protective sheet that shields electronic components from salty reagents. Build-up of salt deposits on unprotected electronic components may combust due to heat generated by electronics when instrument cycles. Completed CAPA identified the root cause as supplier providing defective component (sensor ready valve) and the misalignment of the electronic protective shield during manufacturing of the device.
Action taken
Sysmex sent an Field Correction letter dated April 2017. Product will not be returned, but corrected at the customer site. The correction requires a service visit that will be documented in the Sysmex service Customer Relationship Management (CRM) software. The service report is accepted by the customer at the time of service. Completion of the service records will be tracked by reports generated from the Service CRM. For further Technical Assistance or questions please call (1 888) 879-7639.
Root cause
Component design/selection
Status
Terminated
Product code
GKZ
Quantity affected
3,682 units
Distribution
Nationwide, Canada
Date initiated
2017-03-23
Date posted
2017-04-14
Date terminated
2019-08-22
Location
Lincolnshire, IL

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
SYSMEX XN SERIES (K141681)Sysmex America, Inc.2014-12-05
SYSMEX XN-SERIES (K112605)Sysmex America, Inc.2012-10-19