Syringe, 62RNR 2.5 Microliter Syringe, with needle point style 3 (blunt needle point for use with HPLC injection valves
FDA device recall Z-1169-06 · posted 2006-06-22 · Terminated
Recall details
- Recalling firm
- Hamilton Co
- Reason for recall
- There is an error in the marking of the graduated divisions on syringes in affected lots that shows an amount that is double the actual volume of the syringe
- Action taken
- On 3/28/06, the firm initiated the recall and its notification was via letters explaining the reason for the recall and requesting the product be returned.
- Root cause
- Other
- Status
- Terminated
- Product code
- JQW
- Quantity affected
- 63 units
- Distribution
- Worldwide-The product was distributed to 28 U.S. consignees and 3 foreign consignees. (Canada, Switzerland, Japan)
- Date initiated
- 2006-03-28
- Date posted
- 2006-06-22
- Date terminated
- 2006-12-21
- Location
- Reno, NV