Atlas FDA

Synthes Zero-P VA Implant 7 MM Height Parallel Sterile The Synthes Zero-P VA Implant 7 MM Height Parallel Sterile is a s

FDA device recall Z-0455-2014 · posted 2013-12-05 · Terminated

Recall details

Recalling firm
Synthes USA HQ, Inc.
Reason for recall
A potential non conformance was identified with the Zero-P VA plate where the screw could back out if the blocking mechanism is not visually confirmed to properly cover the head of the screw.
Action taken
Synthes sent an email message dated March 23, 2012, to the sales representative who received the affected product. The email identified the product, the problem, and the action to be taken. The letter requested the sales rep examine inventory, stop using the product, and quarantine the affected product to ensure that it will not be used. Synthes instructed the customer not to remove any product from their account or return any product without further instruction. For questions they were instructed to call 610-719-5364. For questions regarding this recall call 610-719-5000.
Root cause
Process design
Status
Terminated
Product code
OVE
Quantity affected
1
Distribution
Nationwide Distribution - only AZ .
Date initiated
2013-09-30
Date posted
2013-12-05
Date terminated
2015-04-16
Location
West Chester, PA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
SYNTHES ZERO-P VARIABLE ANGLE (VA) (K112068)Synthes Spine Co.Lp2011-11-07