Synthes Titanium Minimally Invasive Reduction Screw System Polyaxial Screw Body, Part Number 04.631.500 Non-cervical spi

FDA device recall Z-0395-2014 · posted 2013-11-26 · Terminated

Recall details

Recalling firm
Synthes USA HQ, Inc.
Reason for recall
This action is being initiated following a detailed investigation in response to reported complaints of post-operative locking cap loosening. Synthes has concluded that if the construct is mis-aligned during final tightening, the potential exists for locking caps to loosen post-operatively in certain situations.
Action taken
Synthes sent an Urgent: Medical Device Product Removal letter on September 7, 2012, to all affected consignees. Consignees were requested to immediately cease using the product, examine their inventory and remove the parts from inventory. If they had the parts, they were further instructed to call Synthes at 1-800-479-6329 to obtain a Return Authorization Number, complete the verification section at the end of the letter and return the verification form. If they did not have the parts, they were requested to complete the verification section at the end of the letter and return the verification form. Consignees with questions were instructed to call 610-719-5450. For questions regarding this recall call 610-719-5000.
Root cause
Use error
Status
Terminated
Product code
NKB
Quantity affected
757
Distribution
Nationwide Distribution including CA, CO, CT, FL, GA, ID, IL, IN, LA, MA, MD, MI, MN, MT, NC, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, and WI.
Date initiated
2012-09-07
Date posted
2013-11-26
Date terminated
2015-12-18
Location
West Chester, PA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
SYNTHES MIRS (K113044)Synthes Spine2011-12-21