Synthes Material Mix (Adjustable Cervical Distractor-Right, Matrix Distractor Rack, T-PAL Spacer Remover, Implant Insert
FDA device recall Z-0393-2014 · posted 2013-11-26 · Terminated
Recall details
- Recalling firm
- Synthes USA HQ, Inc.
- Reason for recall
- One material batch used for production of the affected parts of the Synthes Material Mix were mixed in with an incorrect material with a lower than specified hardness and corrosion resistance.
- Action taken
- SYNTHES sent an Urgent Notice: Medical Device Recall letter dated October 15, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to call Synthes at 1-800-479-6329 to obtain a Return Authorization Number. Complete the Verification Section at the end of the letter by checking the appropriate box indicating affected product was located. Customers were also asked to indicate the number of devices found and note the Return Authorization Number. Return the Verification Form with the product to: Credit/Returns Synthes 1101 Synthes Avenue Monument, CO 80132 Customers should also return the Verification Form if no affected product was found in their inventory. Return all documents to Synthes by fax: 888-679-4875 or Scan/email: FieldAction@synthes.com Customers with questions were instructed to call 610-719-5450 or contact their Synthes Sales Consultant. For questions regarding this recall call 610-719-5000.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- LXH
- Quantity affected
- approximately 44
- Distribution
- Worldwided Distribution: USA (nationwide) including IL, MO, TX, WA and Internationally to Canada and Switzerland.
- Date initiated
- 2013-10-15
- Date posted
- 2013-11-26
- Date terminated
- 2015-08-31
- Location
- West Chester, PA