Synthes Drill Bit for 5.0mm Recon Screws with Large Quick Coupling. This orthopedic manual surgical instrument is a non
FDA device recall Z-0676-2012 · posted 2012-01-11 · Terminated
Recall details
- Recalling firm
- Synthes USA (HQ), Inc.
- Reason for recall
- Synthes is initiating a recall of the Drill Bit for 5.0mm Recon Screws with large Quick Coupling due to the potential for the drill stop to slip on the drill bit.
- Action taken
- On November 2, 2011 the firm sent "Notice: Medical Device Recall" letters to all consignees, with return receipt request. The letter requested consignees cease use of the device and remove the product from inventory. For questions, call 610-719-5450 or contact your sales consultant.
- Root cause
- Component design/selection
- Status
- Terminated
- Product code
- HTW
- Quantity affected
- 577
- Distribution
- Product was distributed nationwide to AL, AK, AZ, AR, CA, CO, CT, FL, GA, HI, ID, IL, IN, IA, KS, KY, ME, MD, MI, MN, MT, LA, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, SD, TX, UT, VA, WA, WV, WI, an
- Date initiated
- 2011-11-02
- Date posted
- 2012-01-11
- Date terminated
- 2015-04-16
- Location
- West Chester, PA