Synthes 2.4mm VA-LCP Two Column Volar Distal Radius Plate and Torque limiting attachment, 1.2Nm Synthes VA-LCP Distal Ra

FDA device recall Z-2057-2013 · posted 2013-08-26 · Terminated

Recall details

Recalling firm
Synthes (USA) Products LLC
Reason for recall
The firm is initiating a Field Safety Correction for the labeling of the 2.4mm VA-LCP Two Column Volar Distal Radius Plate and 1.2 Nm Torque Limiting Attachment. The US Technique guide lists 1.2Nm TLA as an option for use with the 2.4mm Variable Angle LCP (VA-LCP) Two-Column (Standard & Narrow) Volar Distal Radius Plate. The potential exists for the VA locking screw to be driven through the 2.4mm
Action taken
Synthes sent an Urgent Notice: Medical Device Labeling Correction letter dated May 31, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The firm is initiating a Field Safety Correction for the labeling of the 2.4mm VA-LCP Two Column Volar Distal Radius Plate and 1.2 Nm Torque Limiting Attachment. The US Technique guide lists 1.2Nm TLA as an option for use with the 2.4mm Variable Angle LCP (VA-LCP) Two-Column (Standard & Narrow) Volar Distal Radius Plate. The potential exists for the VA locking screw to be driven through the 2.4mm VA-LCP Two Column Volar Distal Radius Plate when used with a 1.2Nm Torque Limiting Attachment. A screw that passes through the VA-LCP plate is not likely to result in a serious adverse event although should it occur, additional medical intervention during the index procedure could be needed. If you have any questions, please call (610) 719-5450, or contact your Synthes sales Consultant.
Root cause
Device Design
Status
Terminated
Product code
HRS
Quantity affected
119, 891
Distribution
Nationwide Distribution including the states of AK. AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MD, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR,
Date initiated
2013-05-31
Date posted
2013-08-26
Date terminated
2015-09-21
Location
West Chester, PA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
SYNTHES (USA) 2.4MM VA-LCP TWO-COLUMN NARROW VOLAR DISTAL RADIUS (K092556)Synthes (Usa)2009-11-17
SYNTHES (USA) 2.4 MM VA-LCP TWO COLUMN VOLAR DISTAL RADIUS PLATES (K083694)Synthes (Usa)2009-03-06
SYNTHES (USA) VARIABLE ANGLE-LOCKING COMPRESSION PLATE (VA-LCP) DISTAL RADIUS SYSTEM (K071184)Synthes (Usa)2007-06-22