FDA device recall Z-2389-2016 · posted 2016-08-11 · Terminated

Recall details

Recalling firm
Synthes (USA) Products LLC
Reason for recall
It was reported that the specified lots of the 2.4mm Titanium Locking Screw Self-Tapping with Stardrive Recess 12mm can break between the head and the shaft. This resulted from the neck feature being too close to the StarDrive feature leaving a thin wall between neck and drive.
Action taken
The firm, DePuy Synthes, sent an "URGENT NOTICE: MEDICAL DEVICE RECALL" letter dated July 14, 2016 to affected customers. The letter described the product, problem and actions to be taken. The customers were instructed to review your inventory and immediately remove and return all unpackaged Locking Screws Self-Tapping with Stardrive Recess 12mm in any affected kits along with any screws still in the original packaging with the part number and lot number identified in the recall notification. Customers were also instructed to call DePuy Synthes at 1-800-479-6329 to obtain a Return Authorization (RA) Number, complete the Verification Section and indicate the number of devices found and return with product to: Credit/Returns, DePuy Synthes, 1101 Synthes Avenue, Monument, CO 80132 by fax to 877-523-9109 or scan/email: Synthes5246@stericycle.com, even if you do not have the product. Note: if the Verification Section is answered on behalf of more than one facility and/or individual, please clearly indicate the name and address of the facility and/or individual. If you have any questions, please call 610-719-5450 or contact your DePuy Synthes Sales Consultant.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
HRS
Quantity affected
41 units
Distribution
US Distribution to states of: AR, CA, CO, DE, FL, GA, IA, IL, KY, LA, MN, NJ, OK, PA, SD, TN, and WA.
Date initiated
2016-06-28
Date posted
2016-08-11
Date terminated
2017-02-02
Location
West Chester, PA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
2.4MM VA-LCP INTERCARPAL FUSION SYSTEM (K103243)Synthes (Usa)2011-01-21
2.4MM VA-LCP DORSAL DISTAL RADIUS PLATES (K102694)Synthes (Usa)2010-12-09
SYNTHES (USA) MODULAR MINI FRAGMENT LCP SYSTEM (K063049)Synthes (Usa)2006-12-18
SYNTHES (USA) LCP MODULAR FOOT PLATES (K050110)Synthes (Usa)2005-03-04
SYNTHES LOCKING DISTAL RADIUS PLATING SYSTEM (K012114)Synthes (Usa)2001-09-28