Syngo.plaza: Syngo.plaza is a Picture Archiving and Communication System (PACS) intended to display, process, read, repo

FDA device recall Z-1027-2015 · posted 2015-01-28 · Terminated

Recall details

Recalling firm
Siemens Medical Solutions USA, Inc
Reason for recall
In case of a system crash, images may not be written to the hard disk and this may result in inconsistencies in the database. In case of a system crash (e.g. blue screen, power outage) images may not be written from cache to the hard disk and might get lost.
Action taken
Siemens sent a Safety Advisory Notice dated January 13, 2015, to all affected customers. The Safety Notice identified the product, the problem, and the action to be taken. Siemens recommended checking patient data in the event of a system crash and advised customers that a software update is planned for mid-2015 that will resolve the issue. For questions regarding this recall call 610-219-6300.
Root cause
Software Manufacturing/Software Deployment
Status
Terminated
Product code
LLZ
Quantity affected
50
Distribution
Nationwide Distribution
Date initiated
2015-01-13
Date posted
2015-01-28
Date terminated
2017-02-16
Location
Malvern, PA