Atlas FDA

SYNGO IMAGING (VERSION V30 and V31); Syngo Imaging is a Picture Archiving and Communication System (PACS) intended to di

FDA device recall Z-1972-2015 · posted 2015-07-02 · Terminated

Recall details

Recalling firm
Siemens Medical Solutions USA, Inc
Reason for recall
During certain clinical workflows safety issues may occur involving monitor assignment order and incorrect display and measurements for images with non square pixel size.
Action taken
Siemens sent a Customer Advisory Notice, dated June 11, 2015, to their customers. The letter identified the affected product, problem and actions to be taken. Customers were also asked to check the software version and consider whether the issue impacts clinical routine and if a software update is necessary.
Root cause
Software design
Status
Terminated
Product code
LLZ
Quantity affected
5
Distribution
US in the states of NC and NE
Date initiated
2015-06-11
Date posted
2015-07-02
Date terminated
2017-03-24
Location
Malvern, PA