syngo Imaging model number 10014063. Affected systems are V30A or earlier with an NFS connected archive and V30A systems
FDA device recall Z-1266-2009 · posted 2009-04-28 · Terminated
Recall details
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Reason for recall
- An issue concerning archived image data that is sporadically saved with corrupted data may occur. This issue may occur when syngo Imaging is configured to archive image data on a network file system (NFS) connected via a connected archive (Mount Point Solution).
- Action taken
- A Customer Safety Advisory Notice, dated 02/09/2009 was sent to affected customers via Update Instruction IM013/09S. The letter informs customers of when the malfunction occurs and the potential issue, provides instructions to avoid its occurrence, and how the issue will be resolved. Customers are to ensure that the Safety Advisory is placed in the system's instructions for use and contact their SIEMENS Uptime Service Senter if they have any questions.
- Root cause
- Software design
- Status
- Terminated
- Product code
- LLZ
- Quantity affected
- 20 units
- Distribution
- Nationwide Distribution -- including states of AL, CA, CT, FL, LA, MI, MO, NC, NJ, OH, PA, TN, and WI.
- Date initiated
- 2009-02-09
- Date posted
- 2009-04-28
- Date terminated
- 2009-09-23
- Location
- Malvern, PA