syngo Dynamics Picture Archiving Communication System The software is a picture archiving and communication system inten

FDA device recall Z-2762-2011 · posted 2011-07-08 · Terminated

Recall details

Recalling firm
Siemens Medical Solutions, USA, Inc
Reason for recall
For systems upgraded to syngo Dynamics version 9.0, if a legacy report generated by version 3.x or earlier is opened in the version 9.0 portal, the legacy report may be deleted and replaced with a new report from the version 9.0 reporting system or a blank report.
Action taken
Siemens Medical Solutions, USA, Inc sent a letter entitled "Customer Safety Advisory Notice" to all direct accounts on May 10, 2011. The letter stated the name and version of the software, problem noted, a work-around to be implemented until a service patch is created to fix the problem, and that the letter should be placed in the software's operator's manual to be seen by all who utilize the software.
Root cause
Software design
Status
Terminated
Product code
LLZ
Quantity affected
38
Distribution
Nationwide Distribution
Date initiated
2011-05-10
Date posted
2011-07-08
Date terminated
2012-07-09
Location
Ann Arbor, MI