SYNCHRON System(s) Lactate (LACT) Reagent. Product Usage: Lactate (LACT) reagent, when used in conjunction with Synchron
FDA device recall Z-1613-2013 · posted 2013-06-27 · Terminated
Recall details
- Recalling firm
- Beckman Coulter Inc.
- Reason for recall
- Beckman Coulter has confirmed customer reports of LACT Lot M210077 failing calibration with error condition OCR LOW. The calibration failures appear to be related to improper shipping or storage conditions (frozen reagent). Failed calibration of LACT may cause a delay in reporting results.
- Action taken
- Beckman Coulter Inc. sent an Urgent Product Correction letter dated May 28, 2013 to affected customers. The letter identified the affected product, the issue, impact, actions to be taken, and resolution. The letter instructed customers to discontinue use of the affected product and request a replacement product. For questions contact Customer Support Center (Call Center) on the Beckman Coulter website at http://www.beckmancoulter.com/customersupport/support/ or call 800-854-3633 in the United States and Canada. Outside the United States or Canada, contact your local Beckman Coulter Representative.
- Root cause
- Under Investigation by firm
- Status
- Terminated
- Product code
- KHP
- Quantity affected
- 3,353 units
- Distribution
- Worldwide Distribution - USA Nationwide and country of Canada.
- Date initiated
- 2013-05-28
- Date posted
- 2013-06-27
- Date terminated
- 2014-06-03
- Location
- Brea, CA