SYMMETRY SURGICAL DISPOSABLE TOWEL CLAMP, REF 19-4137, STERILE EO, Rx only. Apply clamps parallel to the skin with the a

FDA device recall Z-2316-2015 · posted 2015-07-30 · Terminated

Recall details

Recalling firm
Symmetry Surgical, Inc.
Reason for recall
Lack of sterility assurance.
Action taken
The firm initiated their recall by via US certified mail, with return receipt requested, beginning on 06/18/2015. The Urgent Medical Device Recall letter, dated June 16, 2015, identified the affected product, as well as the reason for the recall. The letter discussed the potential harm associated with the issue, and requested that the consignees examine their inventory and discontinue use immediately. Customers should promptly call (800) 251-3000 for a return material authorization (RMA) number to support the return of the impacted lots. Symmetry Surgical Customer Service should be contacted with questions or any other needs.
Root cause
Package design/selection
Status
Terminated
Product code
HXD
Quantity affected
2,708 units
Distribution
Worldwide Distribution -- US, including the states of CA, VA, TX, FL, OH, NY, MS, KY, UT, GA, PA, MO, IN, NV, AL, MI, MA, and LA; and, the countries of Germany and Mexico.
Date initiated
2015-06-16
Date posted
2015-07-30
Date terminated
2016-05-26
Location
Antioch, TN