Symbia T Series SPECT-CT (Single-Photon Emission Computed Tomography and Computed Tomography) System; Siemens Medical So

FDA device recall Z-0584-2010 · posted 2009-12-29 · Terminated

Recall details

Recalling firm
Siemens Medical Solutions USA, Inc.
Reason for recall
Some of the hand controllers of the Symbia S & T systems are missing a resistor switch, and the absence of the switch may lead to unintended system motion during camera set up activities.
Action taken
Siemens issued an Urgent Field Correction Recall letters dated November 18, 2009 and December 11, 2009 to the affected users of the Symbia S and Symbia T Systems, respectivley. The customers were advised of precautionary measures and that Siemens will replace the affected device. They also were advised that a Siemens Customer Service Representative would contact them within a few days to schedule a replacement of the device. For further information, contact Siemens Medical Solutions USA, Inc. at 1-800-888-7436 (USA).
Root cause
Process control
Status
Terminated
Product code
JAK
Quantity affected
97 units
Distribution
Worldwide Distribution -- United States, Algeria, Australia, Austria, Belgium, Canada, Chile, China, Finland, France, Germany, Hong Kong, India, Indonesia, Italy, Korea, Mexico, Netherlands, Norway, P
Date initiated
2009-12-11
Date posted
2009-12-29
Date terminated
2012-08-27
Location
Hoffman Estates, IL