Atlas FDA

Symbia E camera system. Emission Computed Tomography System used to detect or image the distribution of radionuclides in

FDA device recall Z-2043-2014 · posted 2014-07-11 · Terminated

Recall details

Recalling firm
Siemens Medical Solutions USA, Inc.
Reason for recall
Siemens Medical Solutions USA, Inc. has received reports indicating the collimator cart rear casters may become loose.
Action taken
A Customer Safety Advisory Notice dated April 23, 2014 was sent to every direct account. The letter included instructions for direct accounts to discontinue use of the collimator cart and notify the local service representative if they notice or suspect tha one or more casters are loose (which would be indicated by shaking while the cart is in motion). Otherwise, a service representative will be contacting the direct account within the following 3 months to schedule a retention bracket installation. Direct accounts with questions can contact Siemens Medical Solutions USA, Inc. at 800-888-7436.
Root cause
Device Design
Status
Terminated
Product code
KPS
Quantity affected
531 systems
Distribution
Worldwide Distribution- US: Nationwide (AK ; AL ; AR ; AZ ; CA ; CO ; CT ; DE ; FL ; GA ; HI ; IA ; ID ; IL ; IN ; KS ; KY ; LA ; MA ; MD ; ME ; MI ; MN ; MO ; MS ; MT ; NC ; ND ; NE ; NH ; NJ ; NM ;
Date initiated
2014-04-23
Date posted
2014-07-11
Date terminated
2016-12-23
Location
Hoffman Estates, IL

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
SYMBIA (K131634)Siemens Medical Solutions USA, Inc.2013-08-16