Swiss Lithocast Ultra Ultrasound Handpiece (EMS- PN:EL-236) used in combination with Swiss Lithoclast Ultra System .

FDA device recall Z-0684-2007 · posted 2007-03-30 · Terminated

Recall details

Recalling firm
E.M.S. - Electro Medical Systems SA
Reason for recall
Handpiece may become dislodged causing cable damage and possible electric shock.
Action taken
EMS Lyon, France (manufacturer) notified the single US Distributor by letter to contact their accounts and return device for replacement. Boston Scientific, Natick, Ma conducted the sub-recall by letter dated 4/22/05 to all USA accounts.
Root cause
Other
Status
Terminated
Product code
FFK
Quantity affected
160 units
Distribution
Nationwide
Date initiated
2005-03-16
Date posted
2007-03-30
Date terminated
2016-02-02
Location
Nyon, Vaud, Switzerland