SwabCap Disinfecting Cap for Needlefree Connectors: SwabSleeve, REF: SCXT3-5-2000; SwabCap, 200-ct. Box, REF: SCXT3-2000
FDA device recall Z-2116-2025 · posted 2025-07-11 · Open, Classified
Recall details
- Recalling firm
- ICU Medical, Inc.
- Reason for recall
- Due to a manufacturing issue, disinfecting cap for needle-free connectors may have an incomplete seal between the foil lid and plastic container, which may result in isopropyl alcohol evaporation from the sponge, which may result in inadequate disinfection.
- Action taken
- On 6/3/2025, correction notices were mailed and emailed to customers who were asked to do the following: 1) Inform all potential users of the product in your organization of this notification. 2) Discard the affected devices following your institution's process for discarding product. If discarding the device is not immediately possible at your facility, the device should be quarantined until disposal is possible. 3) Complete and return the Customer Response Form to icumedical2818@sedgwick.com 5. DISTRIBUTORS: If you have distributed affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to icumedical2818@sedgwick.com. For further inquiries, please contact firm at the following: Global Complaint Management, globalcomplaints@icumed.com,1-(866)-216-8806; Customer Service, customerservice@icumed.com, 1-(800)-258-5361; Field Action Processing, icumedical2818@sedgwick.com, 1-(866)-328-0114
- Root cause
- Process control
- Status
- Open, Classified
- Product code
- QBP
- Quantity affected
- 1,445,600
- Distribution
- Worldwide - US Nationwide distribution including in the states of TX, SD, ID, LA, IN, CO, OK, MO, MN, OR, AZ, NC, TN, VA, CA, IL, FL, MA, OH, MD, PA, NJ, NY, WA, GA, UT, MI, MS, AR, AL, NV, KS, IA, WV
- Date initiated
- 2025-06-03
- Date posted
- 2025-07-11
- Location
- San Clemente, CA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| SWABCAP AND SWABFLUSH (K130975) | Excelsior Medical Corp. | 2013-12-24 |