SuperElastic Regular force NiTi lower or upper Natural arches. 0.16 size 10 or 100 count packages. Highland Metals Inc.
FDA device recall Z-0922-2014 · posted 2014-02-03 · Terminated
Recall details
- Recalling firm
- Highland Metals, Inc.
- Reason for recall
- Labeling issue: Product labeled as "Upper" may contain "Lower" archwires with a red midline mark. Product labeled ad "Lower" may contain "Upper" archwires with a black midline mark.
- Action taken
- Highland Metals sent an Urgent Field Safety Notice with a response form to all affected consignees on December 13, 2013, via email.. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed to pass the Field Safety Notice to all those who needed to be aware within their organization or to any organization where the potentially affected devices have been transferred. Consignees were instructed to complete the Acknowledgement of Receipt and Action form and fax to 1-408-271-2962 or email to quality@highlandmetals.com. For questions regarding this recall call 408-271-2955.
- Root cause
- Employee error
- Status
- Terminated
- Product code
- DZC
- Quantity affected
- 9 packages of each lot number.
- Distribution
- Worldwide Distribution - USA including IL, KY, MI, NY, CA, FL, LA and Internationally to Switzerland., Denmark, Greece, Australia, England, and Canada.
- Date initiated
- 2013-12-13
- Date posted
- 2014-02-03
- Date terminated
- 2014-02-04
- Location
- San Jose, CA