Sunrise Medical-Hoyer Advance Patient Lifts (Portable and Folding Patient Lift) Sunrise Medical, Inc.
FDA device recall Z-0727-2007 · posted 2007-04-07 · Terminated
Recall details
- Recalling firm
- Joerns Healthcare Inc.
- Reason for recall
- Unapproved Design Change (by base supplier) consisting of a reduction in the number of mast-to-base welds (from four welds to two welds).
- Action taken
- Recall initiated by letter dated 2/21/2007. Sunrise Medical will schedule onsite appointments to complete the base replacement.
- Root cause
- Other
- Status
- Terminated
- Product code
- FSA
- Quantity affected
- 581 units
- Distribution
- Worldwide Distribution --- USA including states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, T
- Date initiated
- 2007-02-21
- Date posted
- 2007-04-07
- Date terminated
- 2010-10-07
- Location
- Stevens Point, WI