Sunquest Encompass The reporting site is using the Encompass software in an acute patient care setting.
FDA device recall Z-2243-2012 · posted 2012-08-22 · Terminated
Recall details
- Recalling firm
- Sunquest Information Systems, Inc.
- Reason for recall
- Sunquest is recalling the Sunquest Encompass versions 2.4 and later because the specimen collect time is displaying and printing one hour later than the actual collect time only when Daylight Saving Time is active.
- Action taken
- The firm, Sunquest, sent a "Urgent-Product Safety Notice: PSN-09-01" letter dated April 13, 2009 all their customers who purchase the Sunquest Encompass versions 2.4 and later. The letter informed the customers of the problem identified and the action to be taken. Customers were instructed to request software correction and to create a service request via client support webpage at www.sunquestinfo.com/supportweb. Then choose Submit a Service Request. Select Issue Type and select Software Request. When Software Request is selected, the Work Queue field automatically fills in with "Software Request." Promptly advise appropriate personnel. Customers with questions can contact Client Support and Services at (877) 239-6337 for US and Canadian sites and at +44(0)0800-960-972 for international sites.
- Root cause
- Software design
- Status
- Terminated
- Product code
- JQP
- Quantity affected
- 23 clients
- Distribution
- Worldwide distribution: USA (nationwide) and country of: Bahamas.
- Date initiated
- 2009-04-13
- Date posted
- 2012-08-22
- Date terminated
- 2012-08-22
- Location
- Tucson, AZ