Atlas FDA

Summit Broach Handle, NON-STERILE, MFG: DEPUY ORTHOPAEDICS, INC., PO BOX 988 WARSAW, IN 46581, 1 800 366 8143 Catalog Nu

FDA device recall Z-0013-2011 · posted 2010-10-06 · Terminated

Recall details

Recalling firm
Depuy Orthopaedics, Inc.
Reason for recall
Firm discovered that some of their straight broach handles are experiencing early failure due to a weakened handle locking mechanism (Leaf Spring).
Action taken
On June 3, 2010, an "Alert: Femoral Broach Handle Repair" went out to the entire sales force/customers via "Sales Mail" and posted to the DePuy Internal website. The Alert stated that "We have discovered that some of our straight broach handles are experiencing early failure due to a weakened handle locking mechanism (Leaf Spring). As you perform routine maintenance on your sets, if you have any of these broaches stamped with the lot numbers listed, it is recommended you have them sent in to be reworked with a new spring device." Send the broach handles to: DePuy Orthopaedics Attn: Repairs 700 Orthopaedic Dr Warsaw, IN 46581 Please include return shipping information If you have any questions regarding this information or need additional information, call 574-372-7333.
Root cause
Process control
Status
Terminated
Product code
LXH
Quantity affected
8358
Distribution
Nationwide distribution: AR, AZ, CA, FL, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, NC, NV, NY, OH, PA, TN, VA, and WI.
Date initiated
2010-06-15
Date posted
2010-10-06
Date terminated
2011-11-30
Location
Warsaw, IN