Suction Instrument with Tubing, Catalog No. 303. Instrument: Yankauer with Vent. *Tubing: 6'' x 1/4'' I.D. Non-Conductiv

FDA device recall Z-0844-04 · posted 2004-07-20 · Terminated

Recall details

Recalling firm
Busse Hospital Disposables
Reason for recall
Firm received complaints of "melting" at tubing connection. The connecting part of the Yankauer becomes soft after extended direct contact with the PVC tubing. This results in a slower rate of suction.
Action taken
Urgent Product Recall Letters and Response Forms were sent to the consignees via Registered Mail on 4/13/04.
Root cause
Other
Status
Terminated
Product code
EBR
Distribution
Distributors and end users (hospitals, fire departments/EMS) located nationwide, in Puerto Rico, St. Thomas, Canada, Dominican Republic, Panama, Uganda, Dubai, and Abu Dhabi, United Arab Emirates.
Date initiated
2004-04-13
Date posted
2004-07-20
Date terminated
2005-02-18
Location
Hauppauge, NY