Suction canisters labels as follows: 1) DeRoyal(R) SUCTION CANISTER, 1000cc, SEMI-RIGID, REF 71-6510, NON-STERILE, Rx On
FDA device recall Z-0097-2015 · posted 2014-10-20 · Terminated
Recall details
- Recalling firm
- DeRoyal Industries Inc
- Reason for recall
- These lots of suction canisters and lids have an insufficient seal under low or intermittent suction condition.
- Action taken
- DeRoyal sent an Urgent Voluntary Recall letter dated August 22, 2014, to all affected consignees by overnight letter or by email. DeRoyal was to follow up with non-responding accounts before October 1, 2014. Customers were instructed to check their inventory for the affected product. If they have affected product, customers were instructed to destory both the liner and the lid and dispose of it according to their facility's procedures. Customers were instructed to return the included Notice of Destruction to recalls@deroyal.com even if they no longer have affected product. Customers with questions were instructed to contact their DeRoyal Sales Representative or DeRoyal Customer Service at 1-800-251-9864. For questions regarding this recall call 865-362-2334.
- Root cause
- Process control
- Status
- Terminated
- Product code
- KDQ
- Quantity affected
- 13,900 units
- Distribution
- Nationwide Distribution including MI, OH, IN, FL, WV, VA, MI, NC, and NY
- Date initiated
- 2014-08-22
- Date posted
- 2014-10-20
- Date terminated
- 2015-04-15
- Location
- Powell, TN