Atlas FDA

Suction canisters labels as follows: 1) DeRoyal(R) SUCTION CANISTER, 1000cc, SEMI-RIGID, REF 71-6510, NON-STERILE, Rx On

FDA device recall Z-0097-2015 · posted 2014-10-20 · Terminated

Recall details

Recalling firm
DeRoyal Industries Inc
Reason for recall
These lots of suction canisters and lids have an insufficient seal under low or intermittent suction condition.
Action taken
DeRoyal sent an Urgent Voluntary Recall letter dated August 22, 2014, to all affected consignees by overnight letter or by email. DeRoyal was to follow up with non-responding accounts before October 1, 2014. Customers were instructed to check their inventory for the affected product. If they have affected product, customers were instructed to destory both the liner and the lid and dispose of it according to their facility's procedures. Customers were instructed to return the included Notice of Destruction to recalls@deroyal.com even if they no longer have affected product. Customers with questions were instructed to contact their DeRoyal Sales Representative or DeRoyal Customer Service at 1-800-251-9864. For questions regarding this recall call 865-362-2334.
Root cause
Process control
Status
Terminated
Product code
KDQ
Quantity affected
13,900 units
Distribution
Nationwide Distribution including MI, OH, IN, FL, WV, VA, MI, NC, and NY
Date initiated
2014-08-22
Date posted
2014-10-20
Date terminated
2015-04-15
Location
Powell, TN

Monitor recalls programmatically