Stryker PCD Precision System 13G Needle with Spineplex, sterile, single use, disposable , 4 units per box. units per box
FDA device recall Z-0862-2012 · posted 2012-01-25 · Terminated
Recall details
- Recalling firm
- Stryker Instruments Div. of Stryker Corporation
- Reason for recall
- Product was packaged with a four inch needle in stead of the correct five inch needle. There is a potential that the needle may not be able to reach the fracture site, which may require additional medical intervention. In addition, if the stylet needs to be changed and the additional stylet that comes in the package is used, this additional stylet is one inch longer than the needle. which could l
- Action taken
- Stryker sent an Urgent Recall Notification letter dated December 7, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. If product was further distributed, customers should forward the recall notification letter and the attached Business Reply Form (BRF) to all affected locations. Immediately check inventory and quarantine any affected product found. Mark product as "RECALLED PRODUCT". Indicate on the BRF the quantity of affected PCD kits they are returning and fill in the form completely. Fax the Business Reply Form to Stryker Instruments Regulatory Department, 866-521-2762. Upon receipt of the BRF, Stryker will contact them to coordinate the return of all affected product that they have on hand. For questions regarding this recall call 269-389-4354. .
- Root cause
- Component change control
- Status
- Terminated
- Product code
- JDZ
- Quantity affected
- 11 kits of the 192 manufactured contain the wrong needle size.
- Distribution
- Nationwide Distribution including FL, PA, NY, OH, IL, KY, MT, WA, CA, MO, HI, IN, SC, MS, WV, KS, TX, and WY
- Date initiated
- 2011-11-30
- Date posted
- 2012-01-25
- Date terminated
- 2012-08-14
- Location
- Portage, MI