Stryker Orthopaedics Triathlon Femoral Distal Augment- Left; 1 single unit per package Howmedica Osteonics Corp. 325 Cor
FDA device recall Z-1370-2013 · posted 2013-05-23 · Terminated
Recall details
- Recalling firm
- Stryker Howmedica Osteonics Corp.
- Reason for recall
- Stryker Orthopaedics received a report that a label on the external packaging (box) of a Triathalon Femoral Distal Augment 10mm- Size 5 Left depicted a glyph of an acetabular shell rather than a femoral distal augment. In addition the label was missing a reference to the "10mm" augment thickness which per the label print is required to appear below the size reference.
- Action taken
- The firm, Stryker Orthopaedics, sent an "URGENT MEDICAL DEVICE REMOVAL" letter dated March 25, 2013 and Product Accountability Forms via Fed Ex to its Consignees/ customers. The letter described the product, problem and actions to be taken. The customers were instructed to complete and return the attached Notification Acknowledgment Form within 5 days via fax to 855-251-3635; contact the hospitals in your territory that have the affected product to arrange return of product; and return affected product to the attention of Regulatory Compliance Stryker Orthopaedics, 325 Corporate Drive, Mahwah, NJ 07430, attach flourescent orange PRODUCT REMEDIATION sticker to your return indicating the PR #RA2013-011 and mark outer box with the words "Product Recall". If you have any questions, feel free to contact Manager, Divisional Regulatory Compliance at (201) 972-2100.
- Root cause
- Software design
- Status
- Terminated
- Product code
- JWH
- Quantity affected
- 11 units
- Distribution
- Nationwide distribution: USA including states of: HI, IN, MS, NC, NJ, OK, TX, and WI.
- Date initiated
- 2013-03-25
- Date posted
- 2013-05-23
- Date terminated
- 2015-05-06
- Location
- Mahwah, NJ
Monitor recalls programmatically