Stryker Orthopaedics Triathlon Femoral Distal Augment- Left; 1 single unit per package Howmedica Osteonics Corp. 325 Cor

FDA device recall Z-1370-2013 · posted 2013-05-23 · Terminated

Recall details

Recalling firm
Stryker Howmedica Osteonics Corp.
Reason for recall
Stryker Orthopaedics received a report that a label on the external packaging (box) of a Triathalon Femoral Distal Augment 10mm- Size 5 Left depicted a glyph of an acetabular shell rather than a femoral distal augment. In addition the label was missing a reference to the "10mm" augment thickness which per the label print is required to appear below the size reference.
Action taken
The firm, Stryker Orthopaedics, sent an "URGENT MEDICAL DEVICE REMOVAL" letter dated March 25, 2013 and Product Accountability Forms via Fed Ex to its Consignees/ customers. The letter described the product, problem and actions to be taken. The customers were instructed to complete and return the attached Notification Acknowledgment Form within 5 days via fax to 855-251-3635; contact the hospitals in your territory that have the affected product to arrange return of product; and return affected product to the attention of Regulatory Compliance Stryker Orthopaedics, 325 Corporate Drive, Mahwah, NJ 07430, attach flourescent orange PRODUCT REMEDIATION sticker to your return indicating the PR #RA2013-011 and mark outer box with the words "Product Recall". If you have any questions, feel free to contact Manager, Divisional Regulatory Compliance at (201) 972-2100.
Root cause
Software design
Status
Terminated
Product code
JWH
Quantity affected
11 units
Distribution
Nationwide distribution: USA including states of: HI, IN, MS, NC, NJ, OK, TX, and WI.
Date initiated
2013-03-25
Date posted
2013-05-23
Date terminated
2015-05-06