Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 64 left, Sterile, Catalog # 504-02-64E-L, UDI (01)07
FDA device recall Z-2094-2020 · posted 2020-05-07 · Terminated
Recall details
- Recalling firm
- Howmedica Osteonics Corp.
- Reason for recall
- The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
- Action taken
- Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
- Root cause
- Packaging
- Status
- Terminated
- Product code
- LPH
- Distribution
- Worldwide distribution including the following states: New York, Arkansas, Oklahoma, Michigan, Delaware, Texas, Louisiana, California, Florida, Arizona, Nevada, Ohio, Colorado, Rhode Island, Iowa, Pen
- Date initiated
- 2020-04-16
- Date posted
- 2020-05-07
- Date terminated
- 2021-11-15
- Location
- Mahwah, NJ
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| Restoration Anatomic Shell (K151264) | Stryker Orthopaedics | 2015-08-10 |