Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Plunger Delivery System, 6", Part Nu

FDA device recall Z-1388-2013 · posted 2013-05-24 · Terminated

Recall details

Recalling firm
Orthovita, Inc., dBA Stryker Orthobiologics.
Reason for recall
There have been a few reported cases of the catheter breaking when removing the Aliquot catheter from the needle during vertebroplasty procedures. This has been attributed to bending of the catheter during the retraction from the needle. It was determined to be use errors. The correction is limited to a revision of the Directions for Use.
Action taken
Stryker sent an Urgent Medical Device Correction letter dated November 10, 2012 to all affected customers. The letter identified the affected product, problem issue, potential hazards and actions to be taken. Customers were instructed to immediately review this notification and the attached IFU. Customers were asked to complete and sign the enclosed Business Reply Form and fax a copy to 1-484-323-8803. For questions contact Stryker Orthobiologics.
Root cause
Labeling design
Status
Terminated
Product code
OAR
Quantity affected
164
Distribution
USA Nationwide Distribution in the states of AZ, CA, CO, CT, DE, FL, GA, HI, IL, IN, KS, KY, LA, MD, ME, MI, MO, NC, ND, NH, NJ, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, and WI.
Date initiated
2012-11-10
Date posted
2013-05-24
Date terminated
2014-10-29
Location
Malvern, PA