Stryker GMRS Proximal Femoral Surgical Protocol; Global Modular Replacement System Proximal Femoral Resection for Large

FDA device recall Z-0668-2008 · posted 2008-04-22 · Terminated

Recall details

Recalling firm
Stryker Howmedica Osteonics Corp.
Reason for recall
The GMRS Proximal Femoral Surgical Protocol states " The GMRS Proximal Femoral components are fully compatabile with all Stryker V40 femoral heads." This statement is incorrect and conflicts with the Alumina Head label and packaging insert.
Action taken
Stryker sent Urgent Product Recall notification letters and Product Accountability forms via FedEx on 12/21/2007, return receipt requested. An Urgent Product Recall letter dated 1/7/2008, was sent to surgeons who are GMRS users, advising that the GMRS Proximal femoral components are not compatible with Alumina ceramic heads.
Root cause
Labeling design
Status
Terminated
Product code
JDI
Quantity affected
5869
Distribution
Worldwide Distribution - USA, Canada, Mexico, China, Australia, Hong Kong, Singapore, Sweden, Germany, Brussels, the Netherlands, Italy, Poland, Dubai, United Kingdom, Spain, and South Africa.
Date initiated
2007-12-18
Date posted
2008-04-22
Date terminated
2008-10-15
Location
Mahwah, NJ