Atlas FDA

Stryker FlowPort II Adapter: Model number: 00CAT00778 The FlowPort II Adapter is intended to connect the FlowPort II Can

FDA device recall Z-1910-2016 · posted 2016-06-03 · Terminated

Recall details

Recalling firm
Stryker Corporation
Reason for recall
Complaints were received for the Stryker FlowPort II Adapter, and investigation found that the scope sat in the adapter 180 degrees in the wrong direction
Action taken
A recall letter was not sent as all affected devices have been returned. A ship hold was placed on the device on October 16, 2015.
Root cause
Device Design
Status
Terminated
Product code
NBH
Quantity affected
21 units
Distribution
GA, UT, MD, NJ, MT
Date initiated
2016-05-04
Date posted
2016-06-03
Date terminated
2016-06-03
Location
San Jose, CA

Monitor recalls programmatically