Stryker FlowPort II Adapter: Model number: 00CAT00778 The FlowPort II Adapter is intended to connect the FlowPort II Can
FDA device recall Z-1910-2016 · posted 2016-06-03 · Terminated
Recall details
- Recalling firm
- Stryker Corporation
- Reason for recall
- Complaints were received for the Stryker FlowPort II Adapter, and investigation found that the scope sat in the adapter 180 degrees in the wrong direction
- Action taken
- A recall letter was not sent as all affected devices have been returned. A ship hold was placed on the device on October 16, 2015.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- NBH
- Quantity affected
- 21 units
- Distribution
- GA, UT, MD, NJ, MT
- Date initiated
- 2016-05-04
- Date posted
- 2016-06-03
- Date terminated
- 2016-06-03
- Location
- San Jose, CA