Stryker Epic II critical care bed, 115v, Stryker Medical, Portage, MI; Model 2030. Beds are A/C powered hospital beds th
FDA device recall Z-2213-2009 · posted 2009-09-18 · Terminated
Recall details
- Recalling firm
- Stryker Medical Div. of Stryker Corporation
- Reason for recall
- The backrest of the bed may not stay in an upright position (may drift down when left raised).
- Action taken
- Stryker Medical issued an "Urgent - Medical Device Correction" letter dated August 10, 2009 informing customers to locate the affected product listed and verify whether it is functioning properly. In the case the device is not properly functioning, remove from service and contact the firm as soon as possible at 1-800-STRYKER, option 3. Customers are also asked to return an enclosed return receipt post card and also forward the notification to users who have loaned or purchased the device. A representative from Stryker field service will be contacting you to perform device corrections shortly. For further information, contact Stryker Medical at 1-269-324-6527.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- FNL
- Quantity affected
- 12,599 total products
- Distribution
- Worldwide Distribution -- United States, Brazil, Canada, Chile, China, India, Korea, Latin America, Mexico, Singapore and Switzerland.
- Date initiated
- 2009-08-10
- Date posted
- 2009-09-18
- Date terminated
- 2011-04-25
- Location
- Portage, MI
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.