Straumann Narrow Neck (NN) Healing Cap with integral occlusal screw Height: 3.4mm Ref: 048.043
FDA device recall Z-0840-06 · posted 2006-05-04 · Terminated
Recall details
- Recalling firm
- Straumann Manufacturing Inc.
- Reason for recall
- The occlusal screw is inserted in the bottom of the healing cap body rather than correctly through the top
- Action taken
- Straumann USA notified accounts by letter dated 3/30/06 via Federal Express. Accounts are requested to return stock.
- Root cause
- Other
- Status
- Terminated
- Product code
- DZE
- Quantity affected
- 103 units
- Distribution
- Nationwide. CA, CO, CT, IA, IL, IN, FL, GA, MA , MD, MS, NC, NY, OH, OK, PA, TX, WA
- Date initiated
- 2006-03-30
- Date posted
- 2006-05-04
- Date terminated
- 2006-06-14
- Location
- Andover, MA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| TITANIUM HEALING CAPS (K960634) | The Straumann Co. | 1996-06-18 |