Straight Inflow/Outflow Patient Line Kit. Model Number: 902-00037. Sterile, single-use, disposable set used with the Hyp

FDA device recall Z-1675-2025 · posted 2025-04-25 · Open, Classified

Recall details

Recalling firm
Belmont Instrument LLC
Reason for recall
Potential crack in the female connector located on the Heat Exchanger in the set, which will lead to a fluid leak through the crack during priming of the rapid infuser.
Action taken
Belmont Medical notified consignees on about 03/31/2025 via "URGENT MEDICAL DEVICE CORECTION" FedEx or emailed letter. Consignees were instructed to locate affected product and to return the customer Response Form indicating if any affected units were on hand. Consignees were also instructed to follow instructions provided in the letter to inspect affected lots and if any units are identified as having cracks or leaks during priming, they should be discarded. Lot number of the set should be noted and reported to Belmont Medical Technologies. Belmont has indicated that they will replace any product found defective or at the request of the user any affected product that has not been used.
Root cause
Process design
Status
Open, Classified
Product code
LGZ
Quantity affected
208 sets
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, CO, CT, DC, IL, IN, KS, KY, MA, ME, MN, NC, ND, NJ, NM, OH, OR, PA, SC, TN, TX, VA, VT, WA, WI, WV and the countries of Canada, Hungary,
Date initiated
2025-03-31
Date posted
2025-04-25
Location
Billerica, MA