Atlas FDA

Stockert Heater-Cooler 3T; Item Number 16-02-8, H3T/240V/60Hz ; 16-02-82, H3T208V/60Hz ; 16-02-85, H3T/120V/60Hz. Device

FDA device recall Z-1186-2016 · posted 2016-03-17 · Terminated

Recall details

Recalling firm
Sorin Group USA, Inc.
Reason for recall
On August 15, 2012, Sorin informed consignees via letter of changes made to create Version 12 of the Heater-Cooler 3T IFU. The revised IFU included enhancements to achieve water quality that meets the standard for German drinking water.
Action taken
Sorin Germany, via Sorin Colorado, distributed an important information letter to its US customers on August 15, 2012 informing them about an update of the cleaning and infection procedure for the U.S. market that were included in Revision 12 of the IFU.
Root cause
Device Design
Status
Terminated
Product code
DWC
Quantity affected
1125 units
Distribution
Worldwide Distribution - US Nationwide, VA/govt/military
Date initiated
2012-08-15
Date posted
2016-03-17
Date terminated
2017-05-26
Location
Arvada, CO

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
STOCKERT HEATER-COOLER SYSTEM 3T (K052601)Sorin Group Deutschland GmbH2006-06-06