Stille Sonesta Gynecology and Urology Table 6302 and 6302U; an electric cystometric table; Stille AB, Solna, Sweden. Use

FDA device recall Z-0565-2010 · posted 2009-12-23 · Terminated

Recall details

Recalling firm
Stille AB
Reason for recall
There is a potential hazard to the patient if the bolts holding the side rail break, causing the accessories attached at the side rails of the tables and chairs to fall off during a medical procedure.
Action taken
Stille sent Field Safety Notices dated 10/30/09 to the direct accounts, informing them that there is a potential hazard with the Sonesta tables if the bolts holding the side rail break, causing the accessories attached at the side rails of the tables and chairs to fall off during a medical procedure. Stille developed a retrofit kit consisting of six side rail bolts and instructions for installing the bolts, which were sent to the end users with the safety notice. The accounts were requested to replace the side rail bolts with the new bolts provided, and to complete and return the enclosed acknowledgement card to Stille by fax or e-mail as soon as possible and no later than 12/11/09. Contact Stille Surgical Inc. at 224-612-5404 for assistance.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
MMZ
Quantity affected
68 units total
Distribution
Worldwide Distribution -- Including countries of USA and Canada.
Date initiated
2009-11-03
Date posted
2009-12-23
Date terminated
2010-12-10
Location
Solna, Sweden