STERRAD NX Sterilizer, Product Code: 10033, Advanced Sterilization Products, Irvine, CA

FDA device recall Z-0848-2008 · posted 2008-02-08 · Terminated

Recall details

Recalling firm
Advanced Sterilization Products
Reason for recall
User Guide Update: based on user reports about contact with residual hydrogen peroxide from instruments, pouches and trays after sterilization and subsequent light colored residue on these devices after sterilization updated user guidance has been issued.
Action taken
Urgent Product Correction letters were sent to customers beginning on November 13, 2007 via UPS overnight. Customers were notified this action was being taken to clarify instructions contained within the STERRAD System User Guides to reinforce appropriate use and to address two separate issues which have arisen from user reports: (1) User reports about contact with residual hydrogen peroxide from instruments, pouches and trays after sterilization. (2) User reports about light-colored residues observed on instruments, trays and pouches after sterilization. The Product Correction Letters instructed customers to refer to the revised excerpt of the User Guide (attached to the letter) for information about instrument preparation prior to sterilization, particularly with regard to the washing, rinsing and thorough drying of instruments prior to sterilization in the STERRAD Sterilization System. If customers have any questions they were instructed to call ASP Customer Care Center to answer additional questions at (888) 783-7723.
Root cause
Employee error
Status
Terminated
Product code
MLR
Quantity affected
9,686 units total for recalls Z-0845-0848-2008
Distribution
Worldwide: USA, France, Germany, Italy, The UK, Switzerland, Benelux, Uruguay, Venezuela, Argentina, Chile, Colombia, Ecuador, Belgium, Puerto Rico, Hong Kong, Indonesia, Mexico, The Philippines, Swit
Date initiated
2007-11-12
Date posted
2008-02-08
Date terminated
2012-03-14
Location
Irvine, CA