STERRAD NX Sterilization System, Product Code: 10033, 10033-002

FDA device recall Z-1197-2017 · posted 2017-02-14 · Terminated

Recall details

Recalling firm
Advanced Sterilization Products
Reason for recall
Advanced Sterilization Products (ASP) has identified that in a rare sequence of events following an external power outage, the system door interlock can be become disengaged during system reboot for a 20-second duration.
Action taken
Advanced Sterilization will send an Field Safety Notice dated January 9, 2017,to inform customers that Advanced Sterilization Products (ASP) has identified that in a rare sequence of events following an external power outage, the system door interlock can be become disengaged during system reboot for a 20-second duration. ASP is sending the letter to remind customers to follow the instructions in the STERRAD NX System Users Guide. Specifically, after cycle initiation, the user should only open the chamber door of the STERRAD NX System when prompted by the Graphical User Interface (GUI) per the Users Guide. Customers are instructed to complete and return the attached Business Reply Card (double-sided) for their acknowledgement within 3 business days of receipt of this notification letter via mail by using the pre-paid postage, by email at ASP5816@stericycle.com, or by fax to Stericycle: 8662297775. For questions regarding the business reply card, customers are instructed to contact Stericycle at (877) 497- 2425 and reference event # 5816. For further questions regarding this recall please call (949) 453-6400,
Root cause
Software design
Status
Terminated
Product code
MLR
Quantity affected
7,694 units total (5,935 units in US)
Distribution
Worldwide Distribution - USA (nationwide) including Puerto Rico and to the countries of : Australia, China, Hong Kong, India, Indonesia, Korea (South), Malaysia, New Zealand, Philippines, Singapore, S
Date initiated
2017-01-09
Date posted
2017-02-14
Date terminated
2017-11-02
Location
Irvine, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
STERRAD NX Sterilizer, STERRAD 100NX Sterilizer (K151725)Advanced Sterilization Products2016-01-13
STERRAD NX Sterilizer, STERRAD 100NX Sterilizer (K142454)Advanced Sterilization Products2015-04-03
STERRAD NX STERILIZER (K042116)Advanced Sterilization Products2005-04-22