Atlas FDA

Sterrad CycleSure Biological Indicator, Product Code 14324. The Sterrad CycleSure Biological Indicator is intended to be

FDA device recall Z-0066-2010 · posted 2009-10-20 · Terminated

Recall details

Recalling firm
Advanced Sterilization Products
Reason for recall
Biological Indicator indicating a positive result does not maintain yellow color for the entire incubation period and some indicators experience media evaporation prior to the end of the labeled maximum incubation time.
Action taken
An Important Product Labeling Correction - New Instructions for Use dated, June 26, 2009, has been proposed to be sent by the firm for Cyclosure Biological Indicators to consignees. The revised instructions reduce the maximum read time for the indicators to 3 days (72 hours). The letter also states why ASP changed the Instructions. Questions are to be directed to ASP Customer Care Center.
Root cause
Device Design
Status
Terminated
Product code
FRC
Distribution
Worldwide Distribution.
Date initiated
2009-06-23
Date posted
2009-10-20
Date terminated
2011-11-29
Location
Irvine, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
STERRAD CYCLESURE BIOLOGICAL INDICATOR (K031226)Advanced Sterilization Products2003-05-02
STERRAD CYCLESURE BIOLOGICAL INDICATOR (K994055)Advanced Sterilization Products2002-02-13