Atlas FDA

STERRAD 100NX Cassettes The STERRAD Sterilization System is a low temperature general purpose sterilizer used to sterili

FDA device recall Z-0044-2014 · posted 2013-10-22 · Terminated

Recall details

Recalling firm
Advanced Sterilization Products
Reason for recall
Advanced Sterilization Products (ASP) is recalling the three lots (13F049, 13F050, 13F051) STERRAD 100NX Cassettes because ASP has determined that the leak indicator on the cassettes is effective for only 12 months of the cassette's labeled 15-month shelf-life.
Action taken
Advanced Sterilization Products (ASP) sent an Urgent Medical Device Recall letter on September 18, 2013, to all affected customers. The letter informed the customers of the problems identified and the actions to be taken. Customers were instructed to examine their inventory and return all affected cassettes using the enclosed prepaid UPS return lave to Stericycle. Customers were asked to count their inventory and record the data on the enclosed Business Reply Card and Packing Slip. Customers were asked to provide the notice to anyone in their facility that needs to be informed. Customers were also asked to maintain a copy of the notice with the affected product. Customers with questions or suspected problems were instructed to contact Stericycle directly at 1-877-257-7119. For questions regarding this recall call 949-453-6400.
Root cause
Under Investigation by firm
Status
Terminated
Product code
MLR
Quantity affected
2,995 units
Distribution
Nationwide Distribution
Date initiated
2013-09-18
Date posted
2013-10-22
Date terminated
2014-01-13
Location
Irvine, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
STERRAD 100NX STERILIZER (K071385)Advanced Sterilization Products2007-11-29