STERIS Boot Assembly, Traction Device, OT 1000, Part Number REF 10009192

FDA device recall Z-1289-2024 · posted 2024-03-12 · Open, Classified

Recall details

Recalling firm
Baxter Healthcare Corporation
Reason for recall
Three lots of the Boot Assembly, Traction Device are being recalled due to due to a customer report of an incorrectly oriented locking pin which caused the boot to detach from the traction system. If an impacted device is used during surgery, the boot may detach from the traction system, potentially leading to patient harm and a delay in surgery.
Action taken
Baxter issued an Urgent Medical Device Recall notice to its sole consignee on 02/14/2024. The notice explained the the problem with the device, potential risk and requested the following actions be taken: Locate, isolate, and cease all use of the affected product and arrange for the return of the affected units calling the Baxter Order Management Team at 800-433-5774, option 2 then 3, between the hours of 8:00 am and 5:00 pm Eastern Time, Monday through Friday, or emailing Baxter at hrc_amatech_orders@baxter.com. If the product was distributed to other facilities or departments within an institution, please forward a copy of this communication to them. Dealers, wholesalers, distributor/resellers, or original equipment manufacturers (OEM) that distributed any affected product to other facilities, are asked to conduct a consumer-level recall of the affected product. For general questions regarding this communication, please contact the Baxter Order Management Team by phone at 800-433-5774, option 2 then option 3, between the hours of 8:00 am and 5:00 pm Eastern Time, Monday through Friday, or by email at hrc_amatech_orders@baxter.com.
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
FWZ
Quantity affected
14 units
Distribution
US Nationwide distribution in the state of Alabama.
Date initiated
2024-02-14
Date posted
2024-03-12
Location
Deerfield, IL