Steris Amsco Sonic Energy Cleaner, Part Nos.: P080000005, B018905020, B018905JBJ, and B018905JBJSS
FDA device recall Z-1407-06 · posted 2006-08-25 · Terminated
Recall details
- Recalling firm
- Steris Corporation
- Reason for recall
- Smoke, sparking and fire hazard-Units manufactured between November 2000 and January 2005 are susceptible to water damage at the lid switch cable wiring harness, which may result in smoking, sparking, and fire.
- Action taken
- The firm sent out a certified letter to all affected consignees alerting them to problems with the device on 6/22/06.
- Root cause
- Other
- Status
- Terminated
- Product code
- FLG
- Quantity affected
- 322 units
- Distribution
- Worldwide USA and countries of Japan, Canada, Hong Kong, Korea, Lebanon, Qatar, Poland, Russia, Saudi Arabia, Singapore, Taiwan, Germany, Northern Mariana Islands, and Malaysia.
- Date initiated
- 2006-06-22
- Date posted
- 2006-08-25
- Date terminated
- 2007-01-08
- Location
- Montgomery, AL