Atlas FDA

Sterile Procedural Trays, labeled as the following: a. BASIN SET-UP PACK b. LAP CHOLE CDS c. MAJOR CDS d. UROLOGY MINOR

FDA device recall Z-1079-2023 · posted 2023-02-06 · Open, Classified

Recall details

Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Reason for recall
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
Action taken
Medline Industries, LP notified consignees via email and first-class mail on 11/15/2022. Consignees were instructed to check stock for affected product and immediately quarantine. For any affected units, customers were instructed to over-label the sterile kits with a warning label indicating "Warning To Avoid Injury, Please Handle Blades with Care." No product is to be returned. Customers were also instructed to complete and return the response form and notify customers if further distributed.
Root cause
Device Design
Status
Open, Classified
Product code
LRO
Quantity affected
62,930 cases (215,456 units)
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Japan, Mexico, Panama, Turks & Caicos, United Arab Emirates.
Date initiated
2022-11-17
Date posted
2023-02-06
Location
Northfield, IL

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.